Leadless Cardiac Pacemaker: Does Anatomical Position at Implant Affect Long-Term Electrical Performance?
A pacemaker (or artificial pacemaker) is
a medical device which uses electrical impulses, delivered by electrodes
contracting the heart muscles, to regulate the beating of the heart. The
permanent leadless cardiac pacemaker (LCP) has been shown to be safe and
effective in human clinical trials. However, there is little information on the
effect of implant location on LCP performance; the aim of this study was to
determine whether anatomic position affected the long-term pacing performance
of the LCP.
In the study, patients
who enrolled in the Leadless II IDE Clinical Trial and had finished 6 months
follow-up (n = 479, mean age: 75.7 ± 11 years old) were selected for the study. The LCP that was implanted (NanostimTM, St. Jude Medical)
is an entirely self-contained, active-fixation, rate-adaptive pacemaker that is
42-mm long with a maximum diameter of 6.00 mm. The LCP was delivered to the
right ventricle through the use of a specially designed delivery system and was
anchored in the right ventricle with the use of a helical screw-in fixation
electrode at the distal end of the device. The implanting investigators
determined the LCP final position under fluoroscope, which was categorized into
three groups: RV apex (RVA, n = 174), RV apical septum (RVAS, n = 101), and RV
septum (RVS, n = 204). Data on capture threshold (at a 0.4 ms pulse width),
R-wave amplitude and impedance were analyzed at implant, hospital discharge and
2 weeks, 6 weeks, 3 months and 6 months post-implant.
The results
indicated that at implant, the mean capture thresholds in the RVA, RVAS and RVS
were 0.77 ± 0.45, 0.81 ± 0.61 and 0.78 ± 0.59 volts, respectively. R-wave
amplitudes were 8.0 ± 3.0 mV, 7.7 ± 2.9 mV and 7.6 ± 2.9 mV, respectively.
Impedance values were 727 ± 311, 765 ± 333, and 677 ± 227 respectively. There
were no differences among the 3 implant locations in capture threshold or
R-wave amplitudes at 6 months (P > 0.06); however, all 3 performance
parameters significantly improved over time (P < 0.001).
In conclusion, the
LCP implant location does not affect capture thresholds or R-wave amplitudes at
6 months, and there is little effect on impedance. Although implant location
does not appear to be a predictor of electrical performance, additional
long-term data will help guide optimal implant location, which would minimize
the risk of perforation or dislodgement. And additional studies with a longer
duration of follow-up are needed to fully evaluate the pacing performance of
the LCP as these devices last for many years.
Full access: http://mrw.so/3J9KoU
Image by Servier Medical Art, from
Flickr-cc.
评论
发表评论