Author(s)
Antar E. Aly, Ibrahim M. Duhaini, Samia M. Manaa, Sayed M. Tarique, Shehim E. Kuniyil, Huda M. Al Naemi
ABSTRACT
Information
about the peak skin dose and Dose Area product (DAP) from percutaneous
transluminal coronary angioplasty (PTCA) and coronary angiography (CA)
was collected from three catheter application rooms. The range of
maximum photon energy was 50 - 125 kVp and the fluoroscopy time was 0.6 -
52 seconds. Values of up to 143 Gy·cm2 for DAP and 0.752 mGy
for cumulative dose (CD) were found in CA procedures. Otherwise the DAP
and CD for PTCA were found to be 143 Gy·cm2 and 2.287 mGy
respectively in 3rd Quartile. The relation between the fluoroscopy time
and the DAP is also considered. Objectives: The objective of this study
is to obtain information about patient peak skin doses (PSD), dose area
product (DAP), Fluoroscopy Time (FT) and Cumulative Dose (CD) from PTCA
and CA which is the most predominant with respect to high skin doses in
addition to other procedures. The aim of this study is also to assess
the radiation dose received by patients undergoing interventional
radiology procedures, by identifying the procedures that deliver the
highest doses. This study is also helpful to establish the reference
dose level for adult patients undergoing interventional procedure, and
to provide recommendations on how to reduce dose on selected procedures
that have been identified to deliver patient dose values near the ICRP
(International Commission on Radiological Protection) threshold values.
Cite this paper
References
Aly,
A. , Duhaini, I. , Manaa, S. , Tarique, S. , Kuniyil, S. and Naemi, H.
(2015) Patient Peak Skin Dose and Dose Area Product from Interventional
Cardiology Procedures. International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 4, 7-12. doi: 10.4236/ijmpcero.2015.41002.
| [1] | Wagner,
L.K., Eifel, P.J. and Geise, R.A. (1994) Potential Biological Effects
Following High X-Ray Dose Interventional Procedures. Journal of Vascular
and Interventional Radiology, 5, 71-84. http://dx.doi.org/10.1016/S1051-0443(94)71456-1 |
| [2] | Koenig,
T.R., Wolff, D., Mettler, F.A. and Wagner, L.K. (2001) Skin Injuries
from Fluoroscopically Guided Procedures: Part 1, Characteristics of
Radiation Injury. American Journal of Roentgenology, 177, 3-11. http://dx.doi.org/10.2214/ajr.177.1.1770003 |
| [3] | Shope, T.B. (1996) Radiation-Induced Skin Injuries from Fluoroscopy. RadioGraphics, 16, 1195-1199. http://dx.doi.org/10.1148/radiographics.16.5.8888398 |
| [4] | Balter, S. (2001) Interventional Fluoroscopy: Physics, Technology, Safety. Wiley-Liss, New York. |
| [5] | Faulkner, K., Broadhead, D.A. and Harrison, R.M. (1999) Patient Dosimetry Measurement Methods. Applied Radiation and Isotopes, 50, 113-123. http://dx.doi.org/10.1016/S0969-8043(98)00031-1 |
| [6] | Fernandez, J.M., Simon, R., Prieto, C., Martinez, D. and Vano, E. (2006) Medical Physics Service. San Carlos University Hospital Version 12, April 2006. |
| [7] | International Electrotechnical Commission (2000) IEC 2000: Particular Requirements for the Safety of X-Ray Equipments for Interventional Procedure. IEC 60601-60602-60603, Geneva. |
| [8] | Stecker, M.S., Balter, S. and Towbin, R.B. (2009) Guideline for Patient Radiation Dose Management. Journal of Vascular and Interventional Radiology, 20, 263-273. |
| [9] | Hansson, B. and Karambatsakidou, A. (2000) Relationships between Entrance Skin Dose, Effective Dose and Dose Area Product for Patients in Diagnostic and Interventional Cardiac Procedures. Radiation Protection Dosimetry, 90, 141-144. http://dx.doi.org/10.1093/oxfordjournals.rpd.a033104 |
| [10] | Tsapaki,
V., Nada, A., Adnan, B., Abdelkader, B., et al. (2009) Radiation
Exposure to Patients during Interventional Procedures in 20 Countries:
Initial IAEA Project Results. American Journal of Roentgenology, 193,
559-569. http://dx.doi.org/10.2214/AJR.08.2115 eww150119lx |
评论
发表评论