Ocular Response Analyzer to Assess Age-Related Biomechanical Characteristics of the Cornea in a Taiwanese Population
Read full paper at:
http://www.scirp.org/journal/PaperInformation.aspx?PaperID=47037
http://www.scirp.org/journal/PaperInformation.aspx?PaperID=47037
Author(s)
Department of Ophthalmology, Taipei Medical University Hospital, Taipei, Taiwan
Department of Ophthalmology, School of Medicine, College of Medicine, Taipei Medical University,
Taipei, Taiwan
College of Medicine, National Taiwan University, Taipei, Taiwan
Department of Ophthalmology, National Taiwan University Hospital, Taipei, Taiwan.
Department of Ophthalmology, Cathay General Hospital, Taipei, Taiwan.
Department of Ophthalmology, Cathay General Hospital, Taipei, Taiwan.
Nobel Eye Clinic, Taipei, Taiwan.
College of Medicine, National Taiwan University, Taipei, Taiwan Department of Ophthalmology, National Taiwan University Hospital, Taipei, Taiwan.
Department of Ophthalmology, Cathay General Hospital, Taipei, Taiwan.
Department of Ophthalmology, Cathay General Hospital, Taipei, Taiwan.
Nobel Eye Clinic, Taipei, Taiwan.
College of Medicine, National Taiwan University, Taipei, Taiwan Department of Ophthalmology, National Taiwan University Hospital, Taipei, Taiwan.
Purpose:
To investigate biomechanical properties of the cornea as measured using a
Reichert ocular response analyzer (ORA) in a Taiwanese population. Methods: Basic
demographic data were recorded for 1875 Taiwanese subjects. Corneal hysteresis
(CH), corneal resistance factor (CRF), central corneal thickness (CCT), Goldmann-equivalent
intraocular pressure (IOPg), and corneal-compensated intraocular pressure
(IOPcc) were measured with an ORA. Descriptive statistics and multiple
regression analyses were used to assess and compare relevant factors of the CRF
and CH. Results: In total, 1817 eyes were included in this study. The mean patient
age was 21.4 ± 17.01 years. The mean (±SD) CH and CRF values were 10.6 ± 1.81
and 10.6 ± 1.99 mmHg, respectively. The mean CCT was 560.4 ± 38.53 μm. CH was correlated
with IOPcc, CRF, and CCT. In the multiple linear-regression model, CH was
associated with and affected by IOPcc, CRF, and CCT. In another multiple linear-regression
model, CRF was associated with and affected by IOPcc, CH, and CCT. Conclusions:
Measuring biomechanical properties by Reichert ORA, we found that CH and CRF
values of a Taiwanese population were similar to those other Asian populations
and other races. Our findings provided fundamental data of CH and CRF in a
Taiwanese (ethic Chinese) and also Asian population.
Cite this paper
Wang, T. , Lee, S. , Hsu, W. , Chang, D. and Wang,
I. (2014) Ocular Response Analyzer to Assess Age-Related Biomechanical
Characteristics of the Cornea in a Taiwanese Population. International Journal of Clinical Medicine, 5, 650-656. doi: 10.4236/ijcm.2014.512089.
| [1] |
Chui, W.S., Lam, A., Chen, D.
and Chiu, R. (2008) The Influence OF Corneal Properties on Rebound
Tonometry. Ophthalmology, 115, 80-84. http://dx.doi.org/10.1016/j.ophtha.2007.03.061 |
| [2] |
Luce, D.A. (2005) Determining in
Vivo Biomechanical Properties of the Cornea with an Ocular Response
Analyzer. Journal of Cataract & Refractive Surgery, 31, 156-162. http://dx.doi.org/10.1016/j.jcrs.2004.10.044 |
| [3] |
Kirwan, C., O’Malley, D. and
O’Keefe, M. (2008) Corneal Hysteresis and Corneal Resistance Factor in
Keratoectasia: Findings Using the Reichert Ocular Response Analyzer.
Ophthalmologica, 222, 334-337. http://dx.doi.org/10.1159/000145333 |
| [4] |
Shah, S., Laiquzzaman, M.,
Cunliffe, I. and Mantry, S. (2006) The Use of the Reichert Ocular
Response Analyser to Establish the Relationship between Ocular
Hysteresis, Corneal Resistance Factor and Central Corneal Thickness in
Normal Eyes. Contact Lens & Anterior Eye, 29, 257-262. http://dx.doi.org/10.1016/j.clae.2006.09.006 |
| [5] |
Kirwan, C., O’Keefe, M. and
Lanigan, B. (2006) Corneal Hysteresis and Intraocular Pressure
Measurement in Children Using the Reichert Ocular Response Analyzer.
American Journal of Ophthalmology, 142, 990-992. http://dx.doi.org/10.1016/j.ajo.2006.07.058 |
| [6] |
Lim, L., Gazzard, G., Chan,
Y.H., Fong, A., Kotecha, A., Sim, E.L., et al. (2008) Cornea
Biomechanical Characteristics and Their Correlates with Refractive Error
in Singaporean Children. Investigative Ophthalmology & Visual
Science, 49, 3852-3857. http://dx.doi.org/10.1167/iovs.07-1670 |
| [7] |
Elsheikh, A., Wang, D., Rama,
P., Campanelli, M. and Garway-Heath, D. (2008) Experimental Assessment
of Human Corneal Hysteresis. Current Eye Research, 33, 205-213. http://dx.doi.org/10.1080/02713680701882519 |
| [8] |
Kotecha, A., Elsheikh, A.,
Roberts, C.R., Zhu, H. and Garway-Heath, D.F. (2006) Corneal Thickness-
and Age-Related Biomechanical Properties of the Cornea Measured with the
Ocular Response Analyzer. Investigative Ophthalmology & Visual
Science, 47, 5337-5347. http://dx.doi.org/10.1167/iovs.06-0557 |
| [9] |
Kirwan, C. and O’Keefe, M.
(2008) Corneal Hysteresis Using the Reichert Ocular Response Analyser:
Findings Pre- and Post-LASIK and LASEK. Acta Ophthalmologica, 86,
215-218. http://dx.doi.org/10.1111/j.1600-0420.2007.01023.x |
| [10] |
Ortiz, D., Pinero, D., Shabayek,
M.H., Arnalich-Montiel, F. and Alio, J.L. (2007) Corneal Biomechanical
Properties in Normal, Post-Laser in Situ Keratomileusis, and Keratoconic
Eyes. Journal of Cataract & Refractive Surgery, 33, 1371-1375. http://dx.doi.org/10.1016/j.jcrs.2007.04.021 |
| [11] |
Kamiya, K., Shimizu, K. and
Ohmoto, F. (2009) Effect of Aging on Corneal Biomechanical Parameters
Using the Ocular Response Analyzer. Journal of Refractive Surgery, 25,
888-893. http://dx.doi.org/10.3928/1081597X-20090917-10 |
| [12] | Narayanaswamy, A., Chung, R.S., Wu, R.Y., Park, J., Wong, W.L., Saw, S.M., et al. (2011) Determinants of Corneal Biomechanical Properties in an adult Chinese Population. Ophthalmology, 118, 1253-1259. |
| [13] |
Pepose, J.S., Feigenbaum, S.K.,
Qazi, M.A., Sanderson, J.P. and Roberts, C.J. (2007) Changes in Corneal
Biomechanics and Intraocular Pressure Following LASIK Using Static,
Dynamic, and Noncontact Tonometry. American Journal of Ophthalmology,
143, 39-47. http://dx.doi.org/10.1016/j.ajo.2006.09.036 |
| [14] |
Lu, F., Xu, S., Qu, J., Shen,
M., Wang, X., Fang, H., et al. (2007) Central Corneal Thickness and
Corneal Hysteresis During Corneal Swelling Induced by Contact Lens Wear
with Eye Closure. American Journal of Ophthalmology, 143, 616-622. http://dx.doi.org/10.1016/j.ajo.2006.12.031 |
| [15] |
Jorge, J.M., Gonzalez-Meijome,
J.M., Queiros, A., Fernandes, P. and Parafita, M.A. (2008) Correlations
between Corneal Biomechanical Properties Measured with the Ocular
Response Analyzer and ICare Rebound Tonometry. Journal of Glaucoma, 17,
442-448. http://dx.doi.org/10.1097/IJG.0b013e31815f52b8 |
| [16] |
Shen, M., Wang, J., Qu, J., Xu,
S., Wang, X., Fang, H., et al. (2008) Diurnal Variation of Ocular
Hysteresis, Corneal Thickness, and Intraocular Pressure. Optometry and
Vision Science, 85, 1185-1192. http://dx.doi.org/10.1097/OPX.0b013e31818e8abe |
| [17] | Avetisov, S.E., Novikov, I.A., Bubnova, I.A., Antonov, A.A. and Siplivyi, V.I. (2010) Determination of Corneal Elasticity Coefficient Using the ORA Database. Journal of Refractive Surgery, 26, 520-524. eww141112lx |